# Minn. Stat. § 153.26 — USE OF NITROUS OXIDE.

- **Citation**: Minn. Stat. § 153.26
- **Node Type**: section
- **Jurisdiction**: Minnesota (state)
- **Code Body**: Minnesota Statutes (2025 edition) (Minn. Stat.)
- **Official Source**: [https://www.revisor.mn.gov/statutes/cite/153.26](https://www.revisor.mn.gov/statutes/cite/153.26)
- **Canonical URL**: [https://corpuslaw.us/codes/minnesota/chapter-153/section-153-26](https://corpuslaw.us/codes/minnesota/chapter-153/section-153-26)
- **Semantic URL**: [https://corpuslaw.us/codes/minnesota/chapter-153/section-153-26](https://corpuslaw.us/codes/minnesota/chapter-153/section-153-26)

## Hierarchy

Chapter 153 — PODIATRY

## Text

153.26. USE OF NITROUS OXIDE.

A podiatrist may use an oxygen-nitrous oxide mixture for sedation or analgesia only under the following conditions:

(1) the patient is not rendered unconscious;

(2) the nitrous oxide concentration is 70 percent of the mixture or less;

(3) the oxygen-nitrous oxide mixture is administered only by nasal inhalator equipped with scavenging system;

(4) an individual other than the attending podiatrist monitors the equipment and the patient during the administration of oxygen-nitrous oxide sedation and is medically directed and supervised by the podiatrist;

(5) the podiatrist and the individual referred to in clause (4) have successfully completed, after December 31, 1986, a course in nitrous oxide sedation approved by a school of podiatric medicine;

(6) the podiatrist and the individual referred to in clause (4) have completed, after December 31, 1986, a course in cardiopulmonary resuscitation;

(7) oxygen-nitrous oxide is not used in conjunction with any other analgesic or sedative; and

(8) the machine used for administering the oxygen-nitrous oxide mixture includes:

(i) an automatic shutoff valve in the event the oxygen concentration is less than three liters per minute;

(ii) an immediately available oxygen flush capability utilizing a positive pressure mask for pure oxygen delivery;

(iii) an automatic control over maximum nitrous oxide concentration which prevents oxygen concentration from falling below 30 percent of the mixture when nitrous oxide flow is maximal; and

(iv) color-coded valves, tanks, and knobs, and pin index and diameter index safety systems.

1990 c 540 s 2