Tolerances for moisture meters.
7 C.F.R. § 801.6
in effect(a) The maintenance tolerances for Motomco 919 moisture meters used in performing official inspection services shall be:
(1) Headquarters standard meters:
Moisture range
Tolerance
Direct comparison
Sample exchange
Low±0.05 percent moisture, mean deviation from National standard moisture meter using Hard Red Winter wheat
Mid±0.05 percent moisture, mean deviation from National standard moisture meter using Hard Red Winter wheat
High±0.05 percent moisture, mean deviation from National standard moisture meter using Hard Red Winter wheat
(2) All other than Headquarters standard meters:
Moisture range
Tolerance
Direct comparison
Sample exchange
Low±0.15 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat±0.20 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat
Mid±0.10 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat±0.15 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat
High±0.15 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat±0.20 percent moisture, mean deviation from standard moisture meter using Hard Red Winter wheat
(b) The maintenance tolerances for GAC 2100 moisture meters used in performing official inspection services shall be:
(1) Headquarters standard meters. By direct comparison using mid-range Hard Red Winter wheat, ±0.05% mean deviation for the average of the Headquarters standard moisture meters.
(2) All other than Headquarters standard meters. By sample exchange using mid-range Hard Red Winter wheat, ±0.15% mean deviation from the standard meter.