§ 175.51 Veterinary Radiography, Dental and Fluoroscopy.

(a) Fixed radiographic installations: equipment. (1) Collimating devices capable of restricting the useful beam to the area of clinical interest must be used. (2) The x-ray films or image receptor used as the recording medium during the x-ray examination must show substantial evidence of cut-off (beam delineation). (3) A device must be provided which terminates the exposure after a preset time interval or exposure. The exposure switch must be of the dead-man type and must be so arranged that it cannot be operated outside a shielded area. (b) Portable or mobile radiographic installations: equipment. (1) Collimating devices capable of restricting the useful beam to the area of clinical interest must be used. (2) The x-ray film or image receptor used as the recording medium during the x-ray examination must show evidence of cut-off (beam delineation). (3) A device must be provided which terminates the exposure after a preset time interval or exposure. (4) A dead-man type of exposure switch must be provided with a cord of sufficient length so that the operator can stand at least 2 meters (6.5 feet) from the animal patient, the x-ray tube and out of the useful beam. (c) Veterinary dental x-ray units. The requirements of 24 RCNY Health Code § 175.49(a) through (r) applies to veterinary dental units with the exception of the ESE requirements of 24 RCNY Health Code § 175.49(q)(3). For hand held dental x-ray units, the same conditions of operation and restrictions on hand held x-ray units (as described in 24 RCNY Health Code § 175.47(j)) apply. (d) Fluoroscopic installations: equipment. (1) Equipment must be so constructed that the entire cross-section of the useful beam is always intercepted by a primary protective barrier regardless of the panel-screen distance. For conventional fluoroscopes, this requirement may be assumed to have been met if, when the collimating system is opened to its fullest extent, an unilluminated margin is left on all edges of the fluorescent screen regardless of the position of the screen during use. Equipment with an image receptor must be so constructed that the useful beam cannot exceed the limits of the input phosphor. (2) The exposure must automatically terminate when the barrier is removed from the useful beam. (3) The fluoroscopic exposure switch must be of the dead-man type. (4) Provision must be made to intercept the scattered x-rays from the undersurface of the tabletop and other structures under the table. (5) The source-panel or source-tabletop distance must always be at least 30 cm (12 in.) and is recommended to be not less than 38 cm (15 in.). (e) Mobile fluoroscopic equipment must meet the following additional requirements: (1) In the absence of a tabletop, a cone or spacer frame must limit the source- to-skin distance to not less than 30 cm (12 in.). (2) Image intensification must always be provided. (3) It must not be possible to operate a machine unless the useful beam is intercepted by the image receptor. (f) Conditions for operation of equipment. (1) Only persons required for the x-ray procedure shall be in the x-ray room during exposures. (2) When an animal patient must be held in position during exposures, mechanical supporting or restraining devices must be used. (3) Animal patients or films or image receptors must be held by an individual only under extreme conditions when clinically necessary. Such individuals must wear protective gloves having at least 0.25 mm lead equivalent, a protective garment of at least 0.25 mm lead equivalent, and must keep all parts of their body out of the useful beam. (4) The exposure of any individual used for holding animals must be monitored. (5) Pregnant women and individuals under 18 years of age must not hold animal patients or films or image receptors under any conditions. (6) Protective garments of at least 0.25 mm lead equivalent must be available and must be worn by the fluoroscopist during every fluoroscopy examination. (Added City Record 4/24/2019, eff. 5/24/2019)

Ask a question about New York law

Corpus searches the New York sections it holds and shows the ones that match, with their citations. Free, no sign-in.