5-6-2: REQUIREMENTS FOR PERMIT:
Mild Hyperbaric Chambers (Chamber) may be exempt from 5-6-1 , and may be permitted in the City only if they have obtained a valid Mild Hyperbaric Therapy Permit, and can meet all of the following: A. The Chamber is a soft-shell chamber. B. An onsite individual is designated as the Hyperbaric Safety Coordinator. C. Anyone operating the Chamber has completed the minimum training requirements. D. There is an Emergency Management Plan in place. E. There are safeguards to ensure continuous supervision while the Chamber is in use. F. Maintain, and have available for Fire Marshall or his designee, records of all scheduled and unscheduled Maintenance. G. Maintain, and have available for Fire Marshall or his designee's, records of all uses and duration of each use of Chamber. H. Provide proof of pressure relief valves with normal working pressure not to exceed 1.4 bar or 20.3 PSI. I. Provide proof the Oxygen concentrations do not exceed twenty three and a half percent (23.5%) or drop below twenty percent (20%). J. Provide proof of a functioning visual or audible communication method within the interior of the Chamber. K. The life span of any Chamber shall not exceed fifteen (15) years or twenty- five thousand (25,000) hours. (Ord. 49-2024, 8-13-2024)